Electronic Pipette
A perfect blend of exceptional performance, superior ergonomics and a super intuitive graphical interface in one pipette, enabling it in understanding user’s requirement and deliver accordingly.
INTUITIVE GRAPHICAL INTERFACE
5 DIFFERENT PIPETTING MODES
STATE-OF-THE-ART STEPPER MOTOR
EFFICIENT BATTERY PERFORMANCE
AUTO CHARGING POD
3 POINT PRECISE AUTO-CALIBRATION
• Intelligent sleep mode for battery conservation, increasing battery efficiency by 40%.
• Calibrated as per ISO 8655-6 standard at an ISO 17025 accredited calibration laboratory for ultimate peace of mind.
• Exceptionally lightweight for effortless and pain-free pipetting even after prolonged use.
• Stores and remembers 3 frequently used modes for quick and hassle – free usage.
• Lightweight.
• 360o rotational manifold to adjust pipetting angle for convenient handling.
• Easy to load tip cones.
• Smooth and leak free operation.
• Magnetic piston for error free pipetting across all channels.
Cat.No. | Description |
DC-S-01 | Acrylic Stand |
DC-CS-01 | Charging Stand Robust and Durable Charging Stand. Charge the pipette when not in use |
ELECTRONIC PIPETTE (SINGLE CHANNEL) | ||||||
Cat. No. | Volume Range [µl] | Test Volume [µl] | Inaccuracy (±) % | Inaccuracy (±) µl % | Imprecision(±) % | Imprecision(±) µl% |
DC-E-10 | 0.5-10 µl | 1 5 10 | 2.5 2 1 | 0.03 0.10 0.10 | 1.8 1 0.6 | 0.02 0.05 0.06 |
DC-E-100 | 5-100 µl | 10 50 100 | 3 1 0.8 | 0.30 0.50 0.80 | 1 0.5 0.2 | 0.10 0.25 0.20 |
DC-E-200 | 10-200 µl | 20 100 200 | 2.5 0.7 0.6 | 0.50 0.70 0.20 | 0.7 0.30 0.20 | 0.14 0.30 0.40 |
DC-E-1000 | 50-1000 µl | 100 500 1000 | 3 1 0.6 | 3.00 5.00 6.00 | 0.6 0.4 0.2 | 0.60 2.00 2.00 |
DC-E-5000 | 250-5000 µl | 500 2500 5000 | 2.4 1.2 0.6 | 12.00 30.00 30.00 | 0.6 0.25 0.2 | 3.00 6.25 10.00 |
ELECTRONIC PIPETTE (8 CHANNEL) | ||||||
Cat. No. | Volume Range [µl] | Test Volume [µl] | Inaccuracy (±) % | Inaccuracy (±) µl % | Imprecision(±) % | Imprecision(±) µl% |
DC-E8-10 | 0.5-10 µl | 1 5 10 | 8 4 2 | 0.08 0.20 0.20 | 5 2 1 | 0.05 0.10 0.10 |
DC-E8-20 | 2-20 µl | 2 10 20 | 7 3 2 | 0.14 0.30 0.40 | 3 2 1.6 | 0.06 0.20 0.32 |
DC-E8-100 | 5-100 µl | 10 50 100 | 3 1 0.8 | 0.30 0.50 0.80 | 2 0.8 0.3 | 0.20 0.40 0.30 |
DC-E8-200 | 10-200 µl | 20 100 200 | 5 1 0.7 | 1.00 1.00 1.40 | 1.4 0.4 0.25 | 0.28 0.40 0.50 |
DC-E8-300 | 15-300 µl | 30 150 300 | 3 1 0.6 | 0.90 1.50 1.80 | 1 0.5 0.3 | 0.30 0.75 0.90 |
ELECTRONIC PIPETTE (12 CHANNEL) | ||||||
Cat. No. | Volume Range [µl] | Test Volume [µl] | Inaccuracy (±) % | Inaccuracy (±) µl % | Imprecision(±) % | Imprecision(±) µl% |
DC-E12-10 | 0.5-10 µl | 1 5 10 | 8 4 2 | 0.08 0.20 0.20 | 5 2 1 | 0.05 0.10 0.10 |
DC-E12-20 | 2-20 µl | 2 10 20 | 7 3 2 | 0.14 0.30 0.40 | 3 2 1.6 | 0.06 0.20 0.32 |
DC-E12-100 | 5-100 µl | 10 50 100 | 3 1 0.8 | 0.30 0.50 0.80 | 2 0.8 0.3 | 0.20 0.40 0.30 |
DC-E12-200 | 10-200 µl | 20 100 200 | 5 1 0.7 | 1.00 1.00 1.40 | 1.4 0.4 0.25 | 0.28 0.40 0.50 |
DC-E12-300 | 15-300 µl | 30 150 300 | 3 1 0.6 | 0.90 1.50 1.80 | 1 0.5 0.3 | 0.30 0.75 0.90 |